Diode Laser Measurement of Line Strengths and Air-Broadening Coefficients of CO2and CO in the 1.57 Μm Region for Combustion Diagnostics

Tingdong Cai,Guangzhen Gao,Xiaoming Gao,Weidong Chen,Gang Liu
DOI: https://doi.org/10.1080/00268970903547934
2010-01-01
Molecular Physics
Abstract:Spectral measurements of two line pairs of CO2 and CO in the temperature range 300–1000 K at 1.573 µm were performed using a fiber-coupled distributed feedback (DFB) diode laser. The two line pairs can be used in a tunable diode laser (TDL) absorption sensor for simultaneously detecting CO2 and CO gas in a single scan of the diode laser. The spectral parameters (line strengths, air-broadening coefficients and the temperature exponent n) of the two pairs are presented. The measured data agree well with existing databases (HITRAN 2004 and HITRAN 2008), the discrepancies being less than 5% for most of the probed transitions. Although the HITRAN database is a useful tool for sensor design, we found that laboratory measurements of the spectroscopic data for the line pair selected for high-temperature sensors are necessary for establishing the uncertainty for accurate measurements.