Highly Sensitive Diode Laser Absorption Measurements of CO 2 Near 1.57 Μm at Room Temperature

Jie Shao,Xiaoming Gao,Weijun Zhang,Yuncong Yuan,Lixin Ning,Ye Yang,Shixing Pei,Wei Huang
2005-01-01
Abstract:The absolute absorption spectrum intensities of carbon dioxide sample have been recorded with a tunable diode laser spectrometer in the spectral range 6350–6364 cm –1 , which is suitable for the in situ sensing of carbon dioxide in the lower stratosphere using a commercial telecommunication -type diode laser. It was found that the typical uncertainty of experimental line intensities is about 1% compared with the values listed in the HITRAN database, which are calculated by direct numerical diagonalization (DND).