Np-Hplc Method for Determining Phytonadione and Its Impurities

Meng Xue,Yun Xue,Chao Yan,Yan Wang
DOI: https://doi.org/10.1080/10826076.2016.1254653
2016-01-01
Abstract:A normal-phase high-performance liquid chromatography method has been developed to separate vitamin cis-phytonadione (cis-K1), trans-phytonadione (trans-K1), trans-epoxyphytonadione (trans-EK1), menadione (K3), cyclo-menadione (CK3), hydroperoxy-phytonadione (K1OOH), and other impurities on a Zorbax RX-SIL column, characterized by very pure porous silica, low acidity, and low metal content. A dual-mode gradient elution was applied involving simultaneous changes in flow rate and mobile-phase composition. The present method was sensitive enough to detect trans-EK1 (limit of detection, LOD: 1.8 ng), K3 (LOD: 0.6 ng), CK3 (LOD: 1.2 ng), and K1OOH (LOD: 2.4 ng). A linear relationship between peak area and concentration was found for four compounds with correlation coefficient r exceeding 0.9994. The impurities (trans-EK1, K3, and K1OOH) were measured accurately (recovery rate 80-120%) and precisely (relative standard deviation less than 2%). Preparing samples under yellow light (520nm), the K1OOH in the three commercial batches of K1 was no higher than 0.10% (w/w) and was measured with a reproducibility of 3.1% and with an intermediate precision of 15.8%. After validation, the developed method was applied for analyzing the impurities occurring in the commercial samples of synthetic K1 and proved to be suitable for routine quality control.
What problem does this paper attempt to address?