HPLC determination of thiamine and its derivatives in human whole blood

FEI Guoqiang,LI Xiaoge,ZHONG Chunjiu
DOI: https://doi.org/10.3969/j.issn.1673-8640.2012.05.001
IF: 1.865
2012-01-01
Laboratory Medicine
Abstract:Objective To establish a stable and rapid method for determining the levels of human whole blood thiamine and its phosphate ester derivatives by high performance liquid chromatography(HPLC).Methods After blood proteins were precipitated by trichloroacetic acid(TCA),thiamine and its phosphate esters were derivatized using potassium ferricyanide into thiochromes.The derivatives of thiamine and it phosphate esters were separated by a reversed-phase HPLC column and measured by fluoroscope.The method was validated by the analysis of linearity,extraction ratio and repeatability.Results The lower limits of quantification of blood thiamine diphosphate(TDP),thiamin monophosphate(TMP) and thiamine determinations were 8.0,1.5 and 2.5nmol/L respectively.The signal to noise ratio(S/N) for determining thiamine and its derivatives was>10.The linear relationship was stable(r>0.99).The recovery rates of TDP,TMP and thiamine were 103.7%,96.8% and 101.7% respectively.The repeatabilities were good.All the coefficients of variation(CV) were <10%.Conclusions The HPLC with thiamine derivatization by potassium ferricyanide for determining thiamine and its phosphate ester derivatives is reliable and can be used for clinical analysis.
What problem does this paper attempt to address?