Comparison of three methods on assay for vitamin A:UV,NP-HPLC and RP-HPLC

Qian Linghui,Liao Jiayu,Li Yani,Mo Xiaofan,Ma Li,Yao Tongwei
DOI: https://doi.org/10.3969/j.issn.1002-4956.2012.05.013
2012-01-01
Abstract:Objective To compare three different methods used in vitamin A determination in soft capsule and to provide reference for the assay of vitamin A and its preparations.Method UV,NP-HPLC and RP-HPLC were chosen in this research.Cyclohexane was used as solvent in the triple wavelength method of UV;Column of NP-HPLC was filled with Hypersil SiO2(250×4.6mm,5μm) and the mobile phase consisted of n-hexane and isopropyl alcohol(99.8∶0.2);Column of RP-HPLC was packed with C18(150×4.6mm,5μm) and the mobile phase was methanol-water(97∶3).Results The content of vitamin A determined by UV was 95.5%,RSD=1.3%(n=5);112.5%,RSD=3.2%(n=5)by NP-HPLC and 102.3%,RSD=1.0%(n=5) by RP-HPLC.Conclusion Both UV and RP-HPLC are convenient measurements for vitamin A;NP-HPLC shows advantages in the test of impurities,while its stability is relatively poor.
What problem does this paper attempt to address?