Comparison of HPLC and amino acid analyzer methods in analyzing 18 different amino acids in total parenteral nutrition injection(25)

Yang Hong-wei
Abstract:Objective:To compare the RP-HPLC and amino acid analyzer(AAA) methods for determining concentrations of amino acids.Method:Eighteen kinds of amino acids in total parenteral nutrition injection(25) were measured by an HPLC and amino acid analyzer.One C18 column was used for HPLC method with phenyl isothiocyanate precolumn derivation,dithio-dipropanoic acid was added as protective agent and cysteine was determined in the meanwhile;an ion exchange chromatography column was used for AAA method with ninhydrin postcolumn derivation,no protective agent was used and cysteine was determined directly.Result:The repeatability of RP-HPLC and AAA methods were less than 1.1% and 3.1%,respectively.The standard deviation of two methods ranged from 0.19%-8.2%.Conclusion:AAA method is more suitable for determining the concentrations of amino acids than HPLC method.
Medicine,Chemistry
What problem does this paper attempt to address?