Pre-column Derivatization-Hplc Method for Determination of 18 Amino Acids in Enteral Nutrition

XU Qian-yu,XU Chen,CHEN Bin,YU Ying-jia,LI Yan,DUAN Geng-li
DOI: https://doi.org/10.19577/j.cnki.issn10074406.2011.06.011
2011-01-01
Abstract:AIM To develop a pre-column derivatization-HPLC method for the determination of proline in enteral nutrition.METHODS An Agilent C_(18) column(150 mm×4.6 mm,5.0μm) was used with a mobile phase consisting of acetonitrile-sodium acetate(pH 4.2,50 mmol·L~(-1)) for gradient elution analysis.The flow rate was 1.0 mL·min~(-1). The column temperature was 30℃.Theλ_(ex) of FLD was 263 nm and theλ_(em) was 315 nm.The derivatization reagent was 9-fluorenylmethyl chloroformate(FMOC-Cl).RESULTS The linear range of proline was 0.1-1 mg·L~(-1)(r= 0.999 6).The range of average recoveries was 109.42%-112.72%.Both the inter-day RSD and the intra-day RSD were less than 7%.The sample solvent was stable in 24 h.CONCLUSION The method is simple,sensitive and specific which can be applied for determination of proline in enteral nutrition and for quality control of the preparation.
What problem does this paper attempt to address?