The Investigation of Long Distance Projection of Transpharyngeal Projection

任文革,张刚,马绪臣,孙广熙
DOI: https://doi.org/10.3969/j.issn.1003-7632.2001.04.017
2001-01-01
Abstract:Objective The condylus can be shown very clearly with the transpharyngeal projection.But the focus-skin distance is only 10cm with the traditional method.which is shorter than the limitation of ICRP and has a higher risk of radiation.In this investigation,we tred to design a fixture device for this projection with 20 cm focus-skin distance and checked the image quality and its repetition.Methods After measuring on the lateral cephalography of 100 cases and calculating the best projecting angle of vertical and horizontal,we designed and made the long distance projection fixture.We applied the new method for 40 films (20 cases) in clinic.The 40 images were compared with other 40 radiograph projected by traditional method.Results The rate of first level image in the traditional method was 65%,while the rate of new method was 72.5%.The results of repetition of the two methods were similar.Conclusion With the new designed fixture device,the transpharyngeal projection could be taken with 20cm focus-skin distance,the new method is useful for clinic and easy to master.