Retrospective Cine ECG-Gated Multislice CT of the Lung:A Comparison with Conventional HRCT

潘自来,张欢,宋琦,凌华威,张绍斌,吴振芳,陈克敏
DOI: https://doi.org/10.3969/j.issn.1001-9324.2003.08.013
2003-01-01
Abstract:Objective To evaluate retrospective cine ECG-gated multislice CT (ECG-MSCT) scanning in reducing cardiac motion artifacts and in improving image quality of the lung. Materials and Methods Conventional HRCT and ECG-MSCT scanning was performed in 50 cases on a GE scanner (Light Speed QX/i). The patients' heart rates ranged from 58 to 88 per minute (mean 72.8±12 beat/min) . Using 10 mm scan interval and 1×1.25 mm collimator, two rotation scans with time overlapping technique for one plane and 10 ~ 15 planes for one person were used. The observation was focused on the pulmonary parenchyma where the pulsation artifacts were most obvious. The scanning data were input to the workstation, where the post-processing of the images was automatically completed. Using blind method, the image quality was evaluated by three radiologists. Artifacts of different kind between conventional HRCT scann and ECG-MSCT scan were compared. Results ECG-MSCT scanning significantly reduced artifacts caused by cardiac motion, which presented as distortion of pulmonary vessels, double-tract sign and blurred cardiac border. And a statistic difference in the presence of artifacts due to cardiac motion existed between ECG-MSCT and conventional CT scanning ( P 0. 01) . No significant difference in eliminating respiratory artifacts was found between two scanning techniques (P0.05). Conclusion ECG-MSCT scanning technique is simple, it can eliminate commonly seen artifacts due to cardiac motion, thus, improve the image quality.