Effect of retrospective electrocardiography-gating on quality of high-resolution computed tomography images of the lung
Zilai Pan,Huan Zhang,Huawei Ling,Bei Ding,Qi Song,Kemin Chen,Hao Jiang
2004-01-01
Abstract:Objective: To investigate the effect of retrospective electrocardiography-gating on the quality of high-resolution computed tomography images of the lung. Patients and methods: Between October 2002 and August 2003, a total of 50 patients underwent high-resolution computed tomography of the lung in a multidetector scanner with and without the use of the retrospectively electrocardiography-gated cine scanner mode. On each image, the presence of cardiac artefact-related star-like features, 'double tram', and blurred cardiac borders was assessed independently by 3 radiologists who were blinded to the method of image acquisition. Results: Retrospectively electrocardiography-gated high-resolution computed tomography of the lung significantly reduced all 3 types of artefacts studied when compared with conventional high-resolution computed tomography. Not only was the frequency of each artefact reduced in the 300 gated scans analysed (p < 0.05), but scores for the overall presence of artefacts also more often indicated the presence of only 1 type or the absence of all 3 types of artefact (p < 0.001). Conclusion: Retrospective electrocardiography-gating improves image quality of high-resolution computed tomography scans of the lung by reducing cardiac motion artefacts that may mimic disease.