Analysis of sulfur-containing odorous gas mixture by capillary gas chromatography

ZHAO Xu-lan,DU Xiao-song,GAO Chao,LONG Yin,CUI Jin,WU Peng-lin
DOI: https://doi.org/10.13595/j.cnki.issn1000-0720.2015.0275
2015-01-01
Abstract:In this paper,a mixture sample of sulfur- containing odorous gas components like methyl mercaptan( MM),dimethylsulfide( DMS),ethanethiol,carbon disulfide( CS2) and dimethyldisulfide( DMDS) were analyzed and detected by an Rt-silica Bond Plot capillary column( silica gel bonded polar group stationary phase) gas chromatography- flame photometric detector( GC- FPD) system. The effects of GC parameters( column injecting pressure and temperature ascending modes) on separating effects were studied. The repeatability for the five components under optimized conditions was tested. The limits of detection( LODs) of these five gases using self- made Tenax tubular concentrator were obtained. The results indicated that the optimum injection column pressure was 48. 26 k Pa and optimum temperature ascending mode was setting from160 ℃ to 220 ℃ increasing by 30 ℃ per minute,which could realize the separation of these five gases completely. The separation time is 3. 850 min and value of the minimum resolution is up to 2. 525. It has good repeatability under optimized conditions with the relative standard deviation( RSD) of retention times and component resolutions less than 1%,and the max RSD of the peak area is 5. 007%. LODs for MM,DMS,ethanethiol,CS2 and DMDS under optimized conditions were 4. 5 × 10- 2mg/m3,3. 7 × 10- 2mg/m3,2. 1 ×10- 2mg/m3,5. 7 × 10- 2mg/m3 and 2. 2 × 10- 2mg/m3 respectively. This method has advantages of rapidness,sensitivity and good repeatability.