Effects of Hydrocarbon Quenching in Gas-Chromatography when Using the Flame Photometric Detector

G. H. Liu,P. R. Fu
DOI: https://doi.org/10.1007/bf02265869
1989-01-01
Chromatographia
Abstract:The long time retardation of the main hydrocarbon peak in the chimney of the flame photometric detector greatly reduces the responses of later-eluting sulfur compounds. In the absence of hydrocarbons in the flame, the slope (s) of the log I vs. log [S] plot (where I is the sulfur response and [S] is the sulfur concentration in the sample) is of the highest value and is constant for all experimental conditions tested. Flame hydrocarbons cause the s value to decrease, and this is dependent on the oxygen to hydrogen ratio in the flame (O/H) and, under certain conditions, also on the sulfur to carbon ratio (S/C) of the sample. The abnormalities observed in the determination of methyl thiol in natural gas are explained on the basis of the present study.
What problem does this paper attempt to address?