Calibration of Button Type Beam Position Monitor for HLS Ⅱ

Yong-liang YANG,Chao-cai CHENG,Bao-gen SUN,Tian-ji MA,Ping LU,Jun-ying ZOU
DOI: https://doi.org/10.7538/yzk.2014.48.03.0541
2014-01-01
Abstract:According to the high precision requirements of the Hefei Light Source major maintenance and reconstruction project , a new button type beam position monitor (BPM) was designed for HLS Ⅱ electron storage ring .The characteristic impedance and distributed capacitance of button pickup were measured by time-domain reflectometer (TDR) ,and scaling coefficient of the new BPM was determined through antenna method .The TDR test results show that the impedance of pickup is 50 Ω which matches the cable impedance ,and the capacitance is about 2.9 pF w hich meets design requirements . Furthermore , the intuitive mapping diagram and calibration factor of beam position were obtained by antenna scanning .Compared with the design value ,the sensitivity deviation of new BPM is less than 0.5% in the vertical direction and less than 2.8% in the horizontal direction .