Note: Durability Analysis of Optical Fiber Hydrogen Sensor Based on Pd-Y Alloy Film.

Peng-cheng Huang,You-ping Chen,Gang Zhang,Han Song,Yi Liu
DOI: https://doi.org/10.1063/1.4941749
IF: 1.6
2016-01-01
Review of Scientific Instruments
Abstract:The Pd-Y alloy sensing film has an excellent property for hydrogen detection, but just for one month, the sensing film's property decreases seriously. To study the failure of the sensing film, the XPS spectra analysis was used to explore the chemical content of the Pd-Y alloy film, and analysis results demonstrate that the yttrium was oxidized. The paper presented that such an oxidized process was the potential reason of the failure of the sensing film. By understanding the reason of the failure of the sensing film better, we could improve the manufacturing process to enhance the property of hydrogen sensor.
What problem does this paper attempt to address?