Vibration Analysis of Atomic Force Microscope Cantilevers in Contact Resonance Force Microscopy Using Timoshenko Beam Model

Xilong Zhou,Pengfei Wen,Faxin Li
DOI: https://doi.org/10.1016/j.camss.2017.09.005
2017-01-01
Abstract:Timoshenko beam model is employed to investigate the vibration of atomic force microscope (AFM) cantilevers in contact resonance force microscopy (CRFM). Characteristic equation with both vertical and lateral tip-sample contact is derived. The contact resonance frequencies (CRFs) obtained by the Timoshenko model are compared with those by the Euler-Bernoulli model. A method is proposed to correct the wave number obtained by the Euler-Bernoulli model. The forced vibration is compared between the two models. Results reveal that the Timoshenko model is superior to the Euler-Bernoulli model in predicting the vibration characteristics for cantilevers’ higher eigenmodes. © 2017 Published by Elsevier Ltd on behalf of Chinese Society of Theoretical and Applied
What problem does this paper attempt to address?