In-situ Measurement of the Dissolved S 2− in Seafloor Diffuse Flow System: Sensor Preparation and Calibration

Ying Ye,Xia Huang,Yi-wen Pan,Chen-hua Han,Wei Zhao
DOI: https://doi.org/10.1631/jzus.a071360
2008-01-01
Abstract:The preparation approach and calibration result of an improved type of ion selective electrode (ISE), which is used to measure the total dissolved S 2− , are introduced in this paper. The improved Ag/Ag 2 S electrode uses silver wire as the substrate, which is surrounded by electric polymer containing superfine silver powder. After the stabilization of the epoxy-resin, Ag 2 S layer was formed by chemical reaction with 0.2 mol/L (NH 4 ) 2 S solution for 5 min. With Ag/AgCl as reference electrode, the Ag/Ag 2 S electrode can be used to measure dissolved S 2− . The correlation between the measured potentials and the logarithm of dissolved S 2− is found to be linear, within range of the concentration of dissolved S 2− from 10 −2 ∼10 −7 mol/L. The slope of the regression line between measured potential and logarithm of dissolved S 2− is about −27.7, which agrees well with the theoretical Nernst value −29.6. Furthermore, the performance of the improved Ag/Ag 2 S electrode, such as the response time, sensitivity and stability, greatly outweighs the conventional Ag/Ag 2 S electrode.