Clinical Evaluation Of Rtog 0813 Dosimetric Criteria For Lung Sabr Using Iplan Monte Carlo Algorithm

damodar pokhrel,r badkul,hong jiang,c estes,junsoo park,f wang,p kumar
DOI: https://doi.org/10.1016/j.ijrobp.2014.05.1864
2014-01-01
Abstract:RTOG lung stereotactic ablative body radiation therapy (SABR) protocols currently require dose to be calculated using tissue heterogeneity corrections. Dosimetric criteria of RTOG 0813 were established based on the results obtained from non-Monte Carlo (MC) algorithms such as superposition/convolutions. Clinically, MC-based algorithms are now routinely used for lung SABR dose calculations, and confirming that MC calculations in lung SABR meet RTOG guidelines is essential. This report evaluates iPlan MC plans for lung SABR patients using dose-volume histogram normalization per current RTOG 0813 compliance criteria. The treatment plans of 10 centrally located lung cancer patients who underwent MC-based lung SABR using heterogeneity corrected X-ray Voxel Monte Carlo (XVMC) algorithm to 60 Gy in 5 fractions for planning target volume (PTV) V100% = 95% were analyzed. Internal target volumes (ITVs) were delineated on maximum intensity projection (MIP) images of 4D-CT scans. PTV (ITV+5 mm margin) volumes ranged from 18.9 to 62.1 cc (mean = 38±13 cc). Organs at risk (OARs) were delineated on average images. Optimal MC SABR plans were generated using a combination of non-coplanar conformal arcs and beams for the Novalis-TX consisting of micro-MLCs and 6 MV-SRS (1000 MU/min) mode. All plans were evaluated using the RTOG 0813 high and intermediate dose spillage criteria: conformity index (CI), ratio of 50% isodose volume to the PTV (R50%), maximum dose 2 cm away from PTV in any direction (D2 cm), and percent of normal lung receiving 20 Gy or more. Other OAR doses were tabulated, including the volume of normal lung receiving 5 Gy (V5), maximum cord dose, dose to <15 cc of heart and dose to <5 cc of esophagus. All MC treatment plans met RTOG 0813 recommendations for D50% and V20. However, four out of ten MC plans had minor deviations in CI, and another four showed deviations in D2 cm. OAR doses, such as maximum cord dose, dose to <15 cc of heart, and dose to <5 cc of esophagus all satisfied RTOG criteria, except for one patient whose tumor was abutting heart. Lung V5 values ranged from 10% to 28%. MC algorithms better predicts dose distributions inside patient anatomy. Results for our XVMC dose calculations indicate that most of the dosimetric guidelines set by RTOG 0813 were met, except for some minor deviations in CI and D2 cm. When using a highly sophisticated MC algorithm such as XVMC, the RTOG 0813 dosimetric compliance criteria may need to be re-visited. To establish new MC based dose parameters, further investigation with a large cohort of patients including peripheral lung tumors is highly recommended.
What problem does this paper attempt to address?