Application of Polycapillary X Ray Lens to Eliminate Both the Effect of X Ray Source Size and Scatter of the Sample in Laboratory Tomography

Xuepeng Sun,Zhiguo Liu,Tianxi Sun,Longtao Yi,Weiyuan Sun,Fangzuo Li,Bowen Jiang,Yongzhong Ma,Xunliang Ding
DOI: https://doi.org/10.3788/col201513.093401
IF: 2.56
2015-01-01
Chinese Optics Letters
Abstract:A tomography device based on a conventional laboratory x ray source, polycapillary parallel x ray lens (PPXRL), and polycapillary collimating x ray lens (PCXRL) is designed. The PPXRL can collect the divergent x ray beam from the source and focus it into a quasi-parallel x ray beam with a divergence of 4.7 mrad. In the center of quasi-parallel x ray beam, there is a plateau region with an average gain in power density of 13.8 and a diameter of 630 μm. The contrast of the image can be improved from 28.9% to 56.0% after adding the PCXRL between the sample and the detector.