A Method for Automatic Three-Dimensional Particle Tracing Under Laboratory Conditions Using Dynamic X-Ray Computed Microtomography

Judith Marie Undine Siebert,Stefan Odenbach
DOI: https://doi.org/10.1007/s11242-024-02086-9
IF: 3.6097
2024-04-24
Transport in Porous Media
Abstract:This paper presents a method for particle tracing in laboratory X-ray micro-computed tomography ( μ CT) using an adjusted Random Sample Consensus (RANSAC) algorithm combined with least squares ellipse fitting (LSF). For method testing, a setup for the investigation of deep bed filtration (DBF) has been used as an example of a complex process that can be elucidated with such a method. Particle tracking with tomography systems requires high-temporal resolution which can only be achieved with synchrotron radiation computer tomography. Therefore, in this work, it has been demonstrated that instead of particle tracking, particle tracing in opaque systems such as DBF can be performed in laboratory μ CT systems. To achieve particle tracing, dynamic μ CT scans with a duration between 30 and 110 s combined with an exposure time of 0.13 s/projection were executed and during the scan time the filtration was performed, causing parabola shaped motion artefacts. The developed method exploits the motion artefacts created by the particle motion during the scan. It could be shown that it is possible to trace particles in complex structures within only one 30 s scan. Furthermore, through trace length and time, it is possible to determine the average velocity. Whereby, the accuracy and limits depend on the particle size, particle velocity/data rate and the X-ray attenuation of particle and medium.
engineering, chemical
What problem does this paper attempt to address?
### Problems Addressed by the Paper The paper aims to address the issue of 3D particle tracking in porous media. Specifically, the authors propose a method for automatically performing 3D particle tracking under laboratory conditions using dynamic X-ray computed microtomography (µCT). This method leverages motion artifacts generated by particle movement during scanning, achieved through an adjusted Random Sample Consensus (RANSAC) algorithm combined with Least Squares Fitting (LSF) of ellipses. ### Background and Motivation Deep Bed Filtration (DBF) is a phenomenon widely present in natural and technical processes, such as riverbank filtration, wastewater treatment, oil extraction, and beer purification. During the DBF process, solid particles are transported and retained as they pass through the filter system. These particles are typically much smaller than the characteristic dimensions of the filter structure, leading to deposition on the internal surfaces at various depths rather than forming a filter cake on top of the filter material. The DBF process is complex, involving multiple transport, attachment, and detachment mechanisms. Although high-resolution synchrotron computed tomography (CT) can be used for single-particle tracking, the availability of synchrotron facilities is limited and costly. Therefore, researchers need a method that enables particle tracking under ordinary laboratory conditions. The method proposed in this paper utilizes motion artifacts generated by particle movement during scanning, achieved through an adjusted RANSAC algorithm and Least Squares Fitting of ellipses, to perform 3D particle tracking in a laboratory µCT system. ### Main Contributions 1. **Development of a New Particle Tracking Method**: This method enables 3D particle tracking using X-ray microtomography (µCT) under laboratory conditions without the need for expensive synchrotron facilities. 2. **Validation of the Method's Effectiveness**: The method's applicability and accuracy were experimentally validated in opaque geometries of varying complexity. 3. **Improved Particle Tracking Efficiency**: The method can track particles in a single 30-second scan and determine the average velocity through tracking length and time. ### Method Overview 1. **Noise Suppression and Binarization**: Preprocessing the images, including Gaussian filtering and binarization, to reduce background noise. 2. **Ellipse Fitting**: Using an adjusted RANSAC algorithm combined with the Least Squares method to fit ellipses and accurately locate particle positions. 3. **Coordinate Determination**: Determining particle coordinates through the vertices of the fitted ellipses. ### Experimental Setup - **Experimental Materials**: Using 3D-printed filter models and experimental conditions, including different single-channel cells and porous systems. - **Particles and Medium**: Using barium titanate particles and glycerol as the medium, as glycerol has high viscosity and X-ray attenuation characteristics similar to photopolymers. - **CT System**: Using a self-built laboratory µCT system, including a nano-focus X-ray tube and a flat-panel detector. ### Conclusion The study successfully developed a method for 3D particle tracking using X-ray microtomography under laboratory conditions, providing a new tool for studying particle movement in porous media. This method is not only suitable for laboratory environments but also effectively handles complex opaque systems.