Remarks on Ion Generation for Cpc Detection Efficiency Studies in Sub-3-Nm Size Range

J. Kangasluoma,H. Junninen,K. Lehtipalo,J. Mikkila,J. Vanhanen,M. Attoui,M. Sipila,D. Worsnop,M. Kulmala,T. Petaja
DOI: https://doi.org/10.1080/02786826.2013.773393
2013-01-01
Aerosol Science and Technology
Abstract:We studied the chemical composition of commonly used condensation particle counter calibration ions with a mass spectrometer and found that in our calibration setup the negatively charged ammonium sulphate, sodium chloride and tungsten oxide are the least contaminated whereas silver on both positive and negative and the three mentioned earlier in positive mode are contaminated with organics. We report cut-off diameters for Airmodus Particle Size Magnifier (PSM) 1.1, 1.3, 1.4, 1.6 and 1.6-1.8 nm for negative sodium chloride, ammonium sulphate, tungsten oxide, silver and positive organics, respectively. To study the effect of sample relative humidity on detection efficiency of the PSM we used different humidities in the differential mobility analyzer sheath flow and found that with increasing relative humidity also the detection efficiency of the PSM increases.