Laboratory Characterization of A Size-Resolved Cpc Battery to Infer the Composition of Freshly Formed Atmospheric Nuclei

Chongai Kuang,Juha Kangasluoma,Daniela Wimmer,Katrianne Lehtipalo,Jian Wang,Markku Kulmala,Tuukka Petaja
DOI: https://doi.org/10.1063/1.4803237
2013-01-01
Abstract:A size-resolved condensation particle counter battery (SR-CPCb) was developed to infer the composition of freshly nucleated aerosol down to 1 nm in diameter. This was accomplished by exploiting the strong dependence of CPC counting efficiency on particle composition and CPC working fluid. The SR-CPCb consists of CPCs that are optimized for sub 2 nm particle detection using the following working fluids: diethylene glycol, water, and butanol. Laboratory characterizations of the SR-CPCb were performed for challenge aerosols of various compositions down to 1 nm in diameter.
What problem does this paper attempt to address?