DOI: https://doi.org/10.1007/978-1-4020-6475-3_43
2007-01-01
Abstract:In this study we present a method how to improve the particle detection efficiency of the condensation particle counter. The limit of the CPC UF-02proto was explored. Our investigations show that the instrument background is 0.002 cm−3 at default regime (temperature difference 32.5°C), and the instrument cut-size (D50%) is 4.4 nm. Increasing the temperature difference, the homogeneous nucleation starts around the temperature difference of 39.0°C. Subsequent temperature difference increasing enhances the number concentration of the nucleated particles. However, the experimental results show that the nucleation process can be prevented by aerosol particles existing in the supersaturated region. When the number concentration of 15 nm silver particles exceeds 4,000 cm−3 then the homogeneous nucleation inside CPC disappears. The detection efficiency as a function of the temperature difference is also investigated. The lowest reliable instrument cut-size D50% was 1.8 ± 0.2 nm when the temperature difference was 44.0°C.