Comparison on the consistency between ROCHE URYSIS-2400 and AIKELAI AX-4030 urine dry-chemistry analyzers

Hui WU,Beili WANG,Wei GUO,Baishen PAN
DOI: https://doi.org/10.3969/j.issn.1673-8640.2014.06.017
IF: 1.865
2014-01-01
Laboratory Medicine
Abstract:Objective To compare the consistency of ROCHE URYSIS-2400 and AIKELAI AX-4030 urine dry-chemistry analyzers.Methods A total of 500 urine specimens were detected by the 2 kinds of urine chemistry analyzers in parallel,and the results were analyzed statistically by positive detection rates and accordance rates.A total of 250 cases were randomly selected for reviewing erythrocytes (ERY)and leulcocytes (LEU)by urine sediment microscopy. Experiments by adding vitamin C were used to evaluate URYSIS-2400 and AX-4030 anti-interference abilities against vitamin C.Results The mean difference percentage of specific gravity (SG)between the 2 analyzers was <1%,and all the general consistency rates of the remaining 9 projects were >90%.The consistency rates of URYSIS-2400 and AX-4030 were 82.4% and 84.4% with urine sediment microscopy on ERY,and 76.0% and 83.2% on LEU. URYSIS-2400 and AX-4030 anti-interference abilities against vitamin C were up to 50mg/dL on glucose (GLU)and ERY.Conclusions There is a good consistency between the results of URYSIS-2400 and AX-4030,but urine chemistry analyzer is used only for screening,and the results should be reviewed by urine sediment microscopy in order to meet clinical needs.