Combination of Urine flow cytometer Sysmex UF-100,test-strip and sediment microscopy in urine erythrocyte,leukocyte and casts analysis

Hu Fei-xue,Zou De-xue,Liu Xiao-yi
2007-01-01
Abstract:Objective: To compare the sensitivity and influencing factors in detecting urine erythrocyte, leukocyte and casts by using Urine flow cytometer Sysmex UF-100, test-strip measurement and sediment microscopy. Methods: Urine specimens of 3000 inpatients were collected and every sample was examined by UF-100, test-strip and microscopy, then comparing the sensitivity of these three urine analysis methods. Results: Analyzed by UF-100, the positive rate of urine erythrocyte, leukocyte and casts were 37.8%, 38.8% and 13.7% respectively; Analyzed by test-strip, the positive rate of urine erythrocyte and leukocyte were 33.4% and 30.3% respectively; Analyzed by sediment microscopy, the positive percentage of urine erythrocyte, leukocyte and casts were 31.5%,35.4% and 4.3% respectively. The positive rate and false positive rate of UF-100 was higher compared with test-strip and sediment microscopy. Conclusions: Sediment microscopy can not be replaced by Urine flow cytometer Sysmex UF-100 and test-strip in urine analysis. Combining these three methods would be optimal to improve the accuracy and precision in urine analysis.