Evaluation of the Sysmex UF-1000i for the diagnosis of urinary tract infection.

Jie Wang,Ying Zhang,DongWen Xu,Weijun Shao,Yuan Lu
DOI: https://doi.org/10.1309/AJCP1GT2JXOCQBCZ
2010-01-01
American Journal of Clinical Pathology
Abstract:Diagnosis of urinary tract infection (UTI) is primarily done by microbilogic culture, which is time-consuming and can produce false-positives and false-negatives. Flow cytometry allows for rapid screening of many samples and eliminates culturing We analyzed the Sysmex UF-1000i (TOA Medical Electronics, Kobe, Japan) for accuracy in identifying RBCs and WBCs, casts, bacteria, and epithelia We also evaluated its precision, linear estimation of results, carryover contamination with manual counting was approximately 95% for RBCs and WBCs, epithelia, and casts Its coefficient of variation for bacteria ranged from 4.7% to 15.2% UF-1000i screening for UTIs exhibited great sensitivity (97%), specificity (79%), positive predictive value (70%), negative predictive value (99%), and accuracy (85%) The negative predicitive value remained high even with complex UTI samples. The Sysmex UF-1000i shows great promise in excluding more than 50% of true-negative samples, improving detection efficiency, and reducing laboratory costs.