Investigation of Mode Stirring with Plate on Platform in a Reverberation Chamber

Madeleine Schilliger Kildal,Xiaoming Chen,Per-Simon Kildal,Jan Carlsson
2015-01-01
Abstract:The accuracy of the reverberation chamber has been investigated when mode stirring is performed with a plate placed on a rotatable platform, i.e., plate-on-platform stirring. The accuracy and scanning volume of the platform plate are then compared to the original mode stirring plates that are moving alongside two walls. The results are also compared to physical theoretical models. The mode stirring with a plate on the platform provides STD of the measurements that are a factor 2 worse than with the original plate and platform stirring of the chamber. Still, this is good enough for accurate measurements above 2 GHz, provided the AUT is also located on the platform.
What problem does this paper attempt to address?