APPLICATION OF Py/FTIR TO PYROLYSIS OF PVC

田原宇,吕永康,谢克昌
DOI: https://doi.org/10.3969/j.issn.0253-2409.2002.06.018
2002-01-01
Journal of Fuel Chemistry and Technology
Abstract:This paper describers the pyrolysic process of PVC by means of CDS2000 pyroprobe with a Fourier Transform infrared spectrometer. The pyrolysis instrument is designed to fit into the sample compartment of a standard FTIR bench, and places the filament of pyrlprobe directly in the light path, thus continuous monitoring of the i.r. spectrum of evolved products. A sample of PVC was heated at 25?℃/min and 20?000?℃/s to a final temperature of 1?000?℃. IR scans were taken during the pyrolysis. The resulting spectrum shows the recognizable pattern of HCl in the gas phase, both P and R branches are present and well resolved. It has been deduced that PVC degrades via side group elimination of HCl from polymer backbone, forming a polyene which fragments further to produce aromatics. By correlating the time with the heating rate, the temperature at which product evolution begins can be determined. Pyrolysis/FTIR can be thus as a useful means of probing the process of pyrolysis.