Transcatheter closure of very large secundum atrial septal defects: long-term follow-up study

wu hong,liu jing,zhao xianxian,liu bo,wang yanyan,zhen xing,sun jing,pan jingwei,guo fuzhi,li weiping,wang ersong,qin yongwen
DOI: https://doi.org/10.1136/heartjnl-2011-300867.602
IF: 5.7
2011-01-01
Heart
Abstract:AimsTo evaluate the efficacy and safety of transcatheter closure of very large secundum atrial septal defects (ASD) with long-term follow-up results.Methods and resultsFrom June, 2001 to June, 2008, 74 patients diagnosed as secundum atrial septal defect with a diameter more than 30 mm were enrolled in the study. Percutaneous closure of ASDs were attempted by using the Domestic Shanghai ASD-O guided by transthoracic echocardiography (TTE). For the group of the 74 patients, 68 of them are adults. The closure was successful in 71/74 patients (96%). The mean diameter of the device was 38.2±2.73 mm. During the follow-up, complete closure could be achieved in 100% of patients. 7 adverse events (9.9%) were reported. Most adverse events were categorised as minor and there was one major adverse events (1.4%).ConclusionTranscatheter device closure is an effective management option for patients with very large ASD. Serious complications were rare.
What problem does this paper attempt to address?