Measurement of the D/H, ¹⁸O/¹⁶O, and ¹⁷O/¹⁶O isotope ratios in water by laser absorption spectroscopy at 2.73 μm.

Tao Wu,Weidong Chen,Eric Fertein,Pascal Masselin,Xiaoming Gao,Weijun Zhang,Yingjian Wang,Johannes Koeth,Daniela Brückner,Xingdao He
DOI: https://doi.org/10.3390/s140509027
IF: 3.9
2014-01-01
Sensors
Abstract:A compact isotope ratio laser spectrometry (IRLS) instrument was developed for simultaneous measurements of the D/H, O-18/O-16 and O-17/O-16 isotope ratios in water by laser absorption spectroscopy at 2.73 mu m. Special attention is paid to the spectral data processing and implementation of a Kalman adaptive filtering to improve the measurement precision. Reduction of up to 3-fold in standard deviation in isotope ratio determination was obtained by the use of a Fourier filtering to remove undulation structure from spectrum baseline. Application of Kalman filtering enables isotope ratio measurement at 1 s time intervals with a precision (<1%) better than that obtained by conventional 30 s averaging, while maintaining a fast system response. The implementation of the filter is described in detail and its effects on the accuracy and the precision of the isotope ratio measurements are investigated.