Development of a quantum cascade laser absorption spectrometer for simultaneous measurement of 13C–18O and 18O–18O clumping in CO2

Scott T. Wieman,Jason Kapit,Anna P. M. Michel,Weifu Guo
DOI: https://doi.org/10.1002/rcm.9836
IF: 2.586
2024-06-20
Rapid Communications in Mass Spectrometry
Abstract:Rationale Dual clumped isotope paleothermometry determines carbonate formation temperatures by measuring the frequency of 13 C– 18 O (∆ 638 ) and 18 O– 18 O (∆ 828 ) pairs in carbonates. It resolves isotopic kinetic biases and thus enables more accurate paleotemperature reconstructions. However, high‐precision measurements of 18 O– 18 O clumping using current techniques requires large sample sizes and long acquisition times. Methods We developed a mid‐infrared isotope ratio laser spectrometer (IRLS) for simultaneous measurement of the isotopologue ratios ∆ 638 and ∆ 828 in gas‐phase carbon dioxide (CO 2 ) at room temperature. Our IRLS uses a single laser scanning from 2290.7 to 2291.1 cm −1 and a 31 m pathlength optical cell, and it simultaneously measures the five isotopologues required for calculating ∆ 638 and ∆ 828 : 16 O 12 C 16 O, 16 O 13 C 16 O, 16 O 12 C 18 O, 16 O 13 C 18 O, and 18 O 12 C 18 O. In addition, our IRLS can measure 16 O 12 C 17 O, enabling ∆ 17 O analysis. Results At ~20°C and a CO 2 pressure of ~2 Torr, our IRLS system achieved precisions of 0.128‰ and 0.140‰ within 20 s for abundances of the clumped isotopologues 16 O 13 C 18 O and 18 O 12 C 18 O, respectively, and precisions of 0.267‰, 0.245‰, and 0.128‰ for 16 O 12 C 16 O, 16 O 13 C 16 O, and 16 O 12 C 18 O. This yielded precisions of 0.348‰ (∆ 638 ) and 0.302‰ (∆ 828 ) within 25 s. Simulated sample–reference switching highlights the potential of our system and the need for further development. Conclusions We demonstrated simultaneous measurements of ∆ 638 and ∆ 828 in CO 2 to precisions of <0.35‰ within 25 s using a room‐temperature, single‐laser IRLS. Future developments on better resolving 16 O 12 C 16 O and 16 O 13 C 16 O peaks and system temperature control could further improve the measurement precision.
spectroscopy,chemistry, analytical,biochemical research methods
What problem does this paper attempt to address?