[Characteristics of Stress Distribution on Bone Adjacent to Implants under Three Different Loading Directions].

Hong-chang Lai,Zhi-yong Zhang,Bao-wei Zhang,Jian-zhong Zhang,Fu-qiang Zhang,Miao Xue
DOI: https://doi.org/10.3969/j.issn.1006-7248.2002.03.020
2002-01-01
Abstract:OBJECTIVE:To evaluate the stress distribution on bone adjacent to an implant under the vertical load, oblique load and horizontal load.METHODS:Obtained three dimensional radiographic data of the jaw by dental CT scan. Based on the scanning sections, the three-dimensional finite element analysis mandible model was established. Loading vertical, oblique and horizontal forces on the implant, calculated the maximum stress, minimum stress and von mises stress.RESULTS:The peak stress occurred at the cervical bone margin adjacent to the implants;the stress concentration under the vertical loading is easily concentrated than that of vertical loading.CONCLUSION:In the clinical designation,not only the force of bite but also the direction of the force should both be emphasised.