Comparison of Stress Distribution at Cervical Bone Between Perfect Implant and Replace Implant in Delayed Loading

CHEN Shu-ping,ZHOU Yan-min,LENG Zhuo,CONG Zhi-qiang,YANG Yan-zhong,WANG Shuai
DOI: https://doi.org/10.3969/j.issn.1671-587x.2008.01.009
2008-01-01
Abstract:Objective To research the stress distribution in cervical bone of Perfect implant and Replace implant in delayed loading.Methods The three-dimensional finite element(FEM)models of the maxilla anterior part were established by CT scanning,video transportation,transformation technology,three-dimensional finite element(FEM)approach,then the stress distribution at alveolar crest in FEM models of Perfect implant and Replace implant was analyzed.Results When Perfect implant was placed in bone and delayedly loaded,the Von Mises stress and compressive stress of alveolar crest were less than Replace implant in the same condition,while the tensile stress was more.Conclusion The stress distribution of cortical alveolar ridge in Perfect implant with delayed-loading is similar to that in Replace implant.Considering these results invitodynamics exclusively,Perfect implant won’t accelerate the bone resorption speed at cervical alveolar ridge.