Capillary Force-Driven Quantitative Plasma Separation Method for Application of Whole Blood Detection Microfluidic Chip
Xiaohua Fang,Cuimin Sun,Peng Dai,Zhaokun Xian,Wenyun Su,Chaowen Zheng,Dong Xing,Xiaotian Xu,Hui You
DOI: https://doi.org/10.3390/mi15050619
IF: 3.4
2024-05-02
Micromachines
Abstract:Separating plasma or serum from blood is essential for precise testing. However, extracting precise plasma quantities outside the laboratory poses challenges. A recent study has introduced a capillary force-driven membrane filtration technique to accurately separate small plasma volumes. This method efficiently isolates 100–200 μL of pure human whole blood with a 48% hematocrit, resulting in 5–30 μL of plasma with less than a 10% margin of error. The entire process is completed within 20 min, offering a simple and cost-effective approach to blood separation. This study has successfully addressed the bottleneck in self-service POCT, ensuring testing accuracy. This innovative method shows promise for clinical diagnostics and point-of-care testing.
chemistry, analytical,nanoscience & nanotechnology,instruments & instrumentation,physics, applied