Comparison of Imaging Quality of 2-Dimensional Double Inversion Recovery Fast Spin Echo Black-Blood Cardiac MR Sequence with Phase Ordering with Automatic Window Selection and Simultaneous Multiple Volume Algorithm Double Navigator-Gated and Breathholding

JU Hai-yue,YANG Li,CHENG Liu-quan,CAI Zu-long
DOI: https://doi.org/10.13929/j.1003-3289.2011.07.016
2011-01-01
Abstract:Objective To assess imaging quality of 2-dimensional double inversion recovery fast spin echo(DIR FSE) black-blood cardiac MR sequence with phase ordering with automatic window selection(PAWS) and simultaneous multiple volume(SMV) algorithm double navigator-gated(NAV) and breathholding,therefore to evaluate the feasibility and value of the former in clinical application.Methods Fifteen healthy adult volunteers underwent 2DNAV DIR FSE black-blood MR with PAWS and SMV algorithm and standard 2D breathhold(2DBH) DIR FSE black-blood MR imaging.The image quality was assessed and compared qualitatively and quantitatively.Results Good consistency(Kappa=0.573,P0.01) of the 2 radiologists' evaluation of 2DNAV DIR FSE image quality was obtained and the correlation of their relatively diagnostic confidence score was statistically significant(r=0.733,P0.01).SNR of left ventricle myocardium and chamber,and the CNR of myocardium/chamber of 2DNAV DIR FSE sequence markedly decreased comparing with those of 2DBH DIR FSE sequence(t=-15.847,-18.399,-14.824,all P0.05).Conclusion PAWS and SMV algorithm 2DNAV DIR FSE black-blood MRI has advantage in cardiac imaging of poor breathhold subjects.However,2DNAV DIR FSE sequence should not be considered as the first choice of cardiac imaging of excellent breathhold subjects.
What problem does this paper attempt to address?