Comparison of 3D delayed enhanced MR sequence with 2D delayed enhanced MR sequence for the detection of nonviable myocardium

GAO Li,WANG Xiao-ying,JIANG Xue-xiang
DOI: https://doi.org/10.3321/j.issn:1003-3289.2006.02.043
2006-01-01
Abstract:Objective To compare one-breath-hold 3D inversion-recovery gradient-echo MR sequence with multiple-breath-hold 2D inversion-recovery gradient-echo MR sequence for the detection of nonviable myocardium. Methods On the basis of a quantitative and qualitative approach, total 34 lesions of myocardial contrast material-enhanced area, as well as the presence and spatial extent of hyperenhancement, were analyzed separately for each MR image obtained with each sequence in the 19 patients with myocardial infarction. Results Findings for total myocardial contrast-enhanced area agreed well between the two sequences. A high level of agreement was also found for the presence of hyperenhancement, while agreement was poor for the transmural extent of hyperenhancement. Conclusion Findings with the one-breath-hold 3D MR sequence allow assessment of nonviable myocardium with good agreement with those with the multiple-breath-hold 2D MR sequence.
What problem does this paper attempt to address?