Abstract:OBJECTIVES The objective of this study was to compare gadobutrol-enhanced gradient-echo sequence (GRE) acquisition with T2-prepared non-contrast-enhanced steady-state free precession (SSFP) in coronary magnetic resonance angiography at 1.5 T. METHODS Twenty-one subjects successfully completed GRE and SSFP acquisition. Signal-to-noise ratio (SNR), contrast-to-noise ratio, image quality, sharpness, visibility, length, and lumen diameter of vessels were analyzed by 2 experienced radiologists. RESULTS The SNR at whole left circumflex artery, left main artery, and proximal left descending artery (LAD) was significantly higher in SSFP acquisition (P < 0.05). The SNR of distal LAD was slightly higher in GRE acquisition (P < 0.05). The contrast-to-noise ratio at distal LAD, proximal and distal RCA were significantly higher with GRE acquisition (P < 0.05). CONCLUSIONS Double-dose gadobutrol-enhanced GRE and unenhanced SSFP coronary magnetic resonance angiography at 1.5 T have their own characteristics, and the combined use of the 2 methods may be taken into consideration.