Bias and linearity measurement of three axis fluxgate magnetometers using high precise electromagnetic devices

Hongfeng Pang,Shitu Luo,Mengchun Pan,Dixiang Chen,Feilu Luo,Qi Zhang
DOI: https://doi.org/10.1109/ASEMD.2011.6145057
2011-01-01
Abstract:Bias error and linearity error influence the measurement precision of three axis fluxgate magnetometers. So, it is important to obtain and calibrate these errors. Bias error and linearity error of a German made DM-060 digital magnetometer are measured by precise electromagnetic devices. Firstly, bias of each axis is obtained by an alternating weak electromagnetic standard equipment. A method by comparing mean absolute error before and after calibration is proposed to demonstrate the bias results credibility. Then, the rotating method is proposed to measure bias by a horizontal barrel shield equipment. Lastly, linearity error of each axis is obtained by a low DC electromagnetic field standard system, and linearity coefficients are calculated by linearity fitting. Experimental results show that bias of X, Y and Z axis is-2.9 nT,-1.4 nT and 3.9 nT by the alternating weak electromagnetic standard equipment, respectively. Bias of each axis changes to-5 nT,-2.5 nT and 10.5 nT by the horizontal barrel shield equipment a year later. Linearity error of X axis, Y axis and Z axis is 3.78-e005, 4.51-e005 and 7.92-e005, respectively. These experimental results show that there is a promising application to use high precise electromagnetic devices for measuring and calibrating bias error and linearity error. © 2011 IEEE.
What problem does this paper attempt to address?