Analysis of Orientation Errors in Triaxial Fluxgate Sensors and Research on Their Calibration Methods

DOI: https://doi.org/10.5194/gi-2024-5
2024-06-07
Geoscientific Instrumentation Methods and Data Systems
Abstract:Three-axis magnetic flux gate sensors are widely used in Chinese geomagnetic observatories, but due to their directional errors, it is necessary to study error correction methods to improve measurement accuracy. Firstly, the mechanism of directional errors produced by three-axis magnetic flux gate sensors is analyzed, followed by the development of measurement tools for conducting directional error measurement experiments on the high-precision three-axis magnetic flux gate sensors of the Chinese FGM-01 series. Experimental results show that correcting the Z-axis and D-axis directional errors is essential. The observation data after error correction, whether in terms of the standard deviation of its all-day baseline values or the relative difference magnitude with the reference instrument, significantly decrease, demonstrating the clear correction effect and proving the effectiveness of this correction method.
geosciences, multidisciplinary,meteorology & atmospheric sciences
What problem does this paper attempt to address?