A disposable cobalt-based phosphate sensor based on screen printing technology

Lei Zhu,Xiaohong Zhou,Hanchang Shi
2010-01-01
Abstract:A cobalt-based phosphate sensor was successfully fabricated by using the screen printing technology for the determination of phosphate concentration in the aqueous solution. The disposable sensor consisted of fully integrated cobalt (Co) electrode, which was a layer of carbon conductive ink (C) physically doped with Co powder, and Ag/AgCl reference electrode. At 25°C and the Co/C ratio of 1:7, the cobalt-based phosphate sensor showed strong stability and phosphate-selective potential response was in the range of 1 × 10 -5 mol/L to 10 -1 mol/L, yielding a detection limit of 5 × 10 -6 mol/L and a slope of -41.63 mV/decade in acidic solution (pH 4.0, potassium hydrogen phthalate buffer) for H 2PO 4 -.The response time of the screen-printed electrode was about 60s. The proposed screen printed sensor also exhibited significant reproducibility with a small repeated sensing deviation (i.e. relative standard deviation (R.S.D.)> 1%) on a single sensor and a small electrode-to-electrode deviation(i.e. R.S.D. > 3.5%).