Study of Fabrication Conditions of ATP Biosensor Based on Screen-Printed Electrode

Qin Zhu,Bo Liang,Yanchuang Pei,Xuesong Ye,Xiao Liang
DOI: https://doi.org/10.1109/icsens.2016.7808858
2016-01-01
Abstract:Adenosine triphosphate (ATP) biosensor has been widely used in biochemistry, molecule biology and food industry in recent years. Meanwhile, screen-printing technology has been applied in electrochemical researches owing to their advanced properties. A dual enzyme ATP biosensor based on co-immobilization of glucose oxidase and hexokinase on screen-printed electrode was developed in this work. The fabrication conditions of platinization and enzyme immobilization, double-layered enzyme electrode structure and magnesium ion concentration during detection were studied. After the optimization, the sensitivity of the ATP biosensor was enhanced up to 20.1 μA mM -1 cm -2 . The results will contribute to the fabrication of high sensitivity ATP biosensor.
What problem does this paper attempt to address?