Mercury Ion Selective Light Addressable Potentiometric Sensor Based on PVC Membrane

LI Yi,ZHANG Zhou-na,CAI Wei,GAO Xiao-ming,WANG Ping
DOI: https://doi.org/10.3785/j.issn.1008-973x.2010.06.033
2010-01-01
Abstract:A novel polyvinyl chloride(PVC) thin-film light addressable potentiometric sensor sensitive to mercury (Ⅱ) ion was tested by gradient concentrations. PVC matrix with 5-amino-1,3,4-thiadiazole-2-thiol was prepared and used as target material. Light addressable potentiometric sensor(LAPS) sensor was chemically modified by silanization. The sensitivities before and after the reaction are 67.299 mV per pH and 4.999 6 mV per pH,the linear correlations are 0.983 1 and 0.987 7,respectively. Once PVC membrane was deposited,Hg2+ sensitive LAPS exhibits a good stability and repetition. The sensitivity of the potentiometric response shows the slope of 15.381 mV per pHg. The detection limit of the sensor was 1.57 μg/L. The response time maintained 2-4 s,and the application range of pH was located between 3-6.
What problem does this paper attempt to address?