Highly sensitive and selective fiber-optic modal interferometric sensor for detecting trace mercury ion in aqueous solution

mingjie yin,bobo gu,jinwen qian,a ping zhang,quanfu an,sailing he
DOI: https://doi.org/10.1039/c2ay05791c
IF: 3.1
2012-01-01
Analytical Methods
Abstract:A novel type of fiber-optic mercury ion (Hg2+) sensor was prepared by coating poly (N-ethyl-4-vinylpyridinium chloride) (P4VP center dot HCl) and poly (sodium-p-styrenesulfonate) (PSS) ultrathin film on the surface of a thin-core fiber modal interferometer (TCFMI) with layer-by-layer (LbL) electrostatic self-assembly method. The fabricated TCFMI Hg2+ sensor exhibits a high selectivity to Hg2+ and its detection limit and response time are 10(-9) Mand 30 s, respectively. The LbL electrostatic self-assembly process is monitored with a quartz crystal microbalance (QCM) and UV-vis spectroscopy for film growth. Atomic force microscopy (AFM) and field emission scanning electron microscopy (FESEM) were used to characterize the morphology of the ultrathin film. Both demonstrated nanoporous film, which is helpful in shortening the response time of the TCFMI Hg2+ sensor.
What problem does this paper attempt to address?