Device for temperature measurement in counter-current gas-liquid spraying field

Si-yu JIN,Yi-xing LI,Shi-lie WENG
DOI: https://doi.org/10.3969/j.issn.1672-9897.2008.04.014
2008-01-01
Abstract:A new thermocouple temperature measurement instrument for counter-current gas-liquid spraying field was designed. The gas and liquid phase temperature could be measured simultaneously with this device. With the on-line monitoring particle measurement equipment based on the principle of light fluctuation, the separation effectiveness of gaseous phase from the mixed two-phase flow was evaluated at two different situations of vacuum pump and no vacuum pump. By analyzing the diagrams of received light intensity, the results show that copper mesh at the bottom of the instrument can prevent liquid droplets from getting into the shield effectively. If adopting copper mesh with proper mesh size and vacuum pump, there is no signal of particle light fluctuation inside the shield, which means the separation effectiveness of gas-liquid phases is satisfied. So the gaseous phase temperature can be measured accurately and the liquid phase temperature also can be obtained directly with the thermocouple on the top of the shield.
What problem does this paper attempt to address?