Research on pressure sensor location in differential vortex mass flowmeter

Yongmei Huang,Zhiqiang Sun,Hongjian Zhang
DOI: https://doi.org/10.3321/j.issn:0254-3087.2007.09.027
2007-01-01
Abstract:In vortex flowmeter, fluid passes a bluff body and lead to a pressure drop and pressure fluctuation generated by vortex shedding. A new method to measure mass flow rate is proposed, which uses single differential pressure sensor to detect the pressure double change according to the characteristics of flow field in vortex flowmeter. The paper mainly focuses on the research of the pressure sensor location. Experiments were carried out using two kinds of fluid mediums-gas and water; and the relationships between differential pressure signal and flow rate in different locations were acquired. The suitable location for differential pressure sensor is determined to measure mass flow rate accurately. Experimental results manifest that this method is feasible and effective for mass flow rate measurement.
What problem does this paper attempt to address?