THE BEAM CHARACTERISTICS OF INTENSITY-MODULATED RADIOTHERAPY 6MEV STANDING WAVE ACCELERATING TUBE
H.Chen,Qingxiu Jin,Renkai Li,Yu-Chi Lin
2010-01-01
Abstract:The method of intensity-modulated radiotherapy (IMRT) is increasingly concerned by the medical world in recent years. Based on the performance characteristic of IMRT accelerator, a 6MeV S-band on axis-coupled SW, suitable for IMRT, electron linear accelerating tube has been developed in Accelerator Lab of Tsinghua University. This paper provides the design performance characteristics of the tube and the results of the highpower tests, analyzes the performance and problems in the operation. INTRODUCTION Accelerator Lab of Tsinghua University has been researching the axis-coupled SW accelerating structure since 1979. We have succeeded in designing and producing more than 200 medical electron linear accelerators with beam energy of 6MeV, 14MeV and 20MeV, most of which has been provided for many manufactures and institutes. With the development of radiation therapy, IMRT has become an important field and relevant accelerator, in which a maximum dose rate of 600MU/min X-ray (in a 40 ×40 cm field) could be produced, is required. However, it is a great challenge to achieve 600MU/min at 6MeV beam energy in a 40×40 cm field. The current maximum dose rate generated by domestic medical accelerator is only 400MU/min when beam energy and square radiation fields were scaled to the same. We decided to attack this problem from two aspects: • Based on the former 6MeV SW accelerator in our lab, optimizing the structure geometry without increasing the tube length to enhance maximum beam intensity • Developing the magnetron to increase input power into the accelerating structure from 2.6MW (MG5193) to 3.1MW (MG5349). Beside, the effect of target geometry and material should also be considered to reach the desired criteria. Based on the performance characteristic of IMRT accelerator, a 6MeV S-band test structure accelerating tube has been developed. Now the test of operating character is still underway. This report will give the current result. DESIGN PARAMETERS The medical accelerator needs to achieve certain beam energy in limited distance. After making the balance between the accelerating tube length and beam energy, the goal of beam energy is set to 6.3MeV. The required average beam current is 138uA for IMRT. Both the cell structure and beam dynamics have been optimized to maximize the beam intensity by using a simulation code developed in our lab, and a better design has been achieved. The basic operation parameters of the accelerator structure are listed in Table1: Table 1: Parameters of the Structures Parameter: Design value: Accelerator Length: L[mm] 285 Beam Intensity: Im[mA] 130 Average Current I [uA] 138 Beam Energy: P0 [MeV] 6.3 Shunt Impedance: ZT[MΩ/m] 122 Input Power: Pin[MW] 2.4 Input Coupling: β 1.6 One test structure based on the design mentioned above has been fabricated. For the purpose of beam characterization, an 80μm-think titanium vacuum window was welded at the beam exit instead of a target, and the beam could pass through the window. HIGH-POWER RF TESTING Generally, our High-Power RF Test Facility can be divided into three parts: • High-Power RF system which supplies the required input power. • A magnetic analyzer designed to measure the energy spectrum. • A Faraday Cup used to measure the beam intensity. The energy spectrum measurement was performed as follows: the accelerated beam passed through the titanium window and entered into the magnetic analyzer, whose vacuity was kept at 1.5 × 10Pa by the turbine molecular pumps. The electron beam turned about 60 degrees in the analyzer and passed through a 50μm-think titanium window, behind which there was a phosphorescent screen and a CCD camera. Eventually the beam hit the screen and its position could be accurately determined through the observed bright spot. The PC controlled the camera, recorded and analysed the data to get the energy spectrum automatically. When the magnetic analyzer was not operating, the beam would not bend in the magnetic dipole and directly entered into the Faraday Cup where the beam intensity could be measured. Proceedings of IPAC’10, Kyoto, Japan MOPEA016 08 Applications of Accelerators, Technology Transfer and Industrial Relations U01 Medical Applications 97 High power RF system mainly consists of MG5349 magnetron made by E2V Company, four-port circulator made by China and spectrum analyzer utilized for realtime performance monitoring. However, the magnetron has broken down just before the experiment. The former MG5193 2.6MW magnetron was installed to supply the power instead. The details of the magnetic analyzer and auxiliary equipment are given in Fig.1. A very stable (better than 0.1%) and adjustable power system supplies the power to the magnetic analyzer. Magnetic shielding has been installed at the entrance and exit to reduce the fringe field effect. Besides, in order to enhance the resolution, there is a 38mm-long and 0.2mm wide slit at the entrance to limit the beam size. And accurate discrete calculations were performed in the computer with the carefully measured data of the magnetic field. Figure 1: The magnetic analyzer and auxiliary equipment. Taking all various factors into consideration, the energy resolution can be promoted to less than 1%. EXPERIMENTAL RESULT The high power test has been performed on our system described above. The test result will be presented as follows: