Restraining speckle of laser imaging system in accurate measurement

Shu-Yan REN,Zhuo ZHANG,Guo-Dong LIU,Zhi-Tao ZHUANG,Zhao-Bang PU
DOI: https://doi.org/10.3321/j.issn:1004-924X.2007.03.006
2007-01-01
Abstract:Three factors of restraining speckle noise in detection system,coherence,phase difference and polarization of the incident ray,were analyzed in detail, and the relations between these factors and speckle contrast were also presented.Calculation results show that when coherence length is less than 1 mm,speckle contrast is reduced to 0.5,which is advantageous to obtaining excellent measuring accuracy and keeping good details for a image. On the basis of theoretical analysis,the solution and optimization to reduce speckle was proposed to restrain speckle noise,improve illumination quality and enhance accuracy of measurement.
What problem does this paper attempt to address?