Abstract:A novel chemiluminescence (CL) flow sensor, based on a plant tissue reactor, for the determination of hydrogen peroxide was proposed. It was prepared by using a potato tissue column as molecular recognition element. Luminol was oxidized by hydrogen peroxide under the catalysis of peroxidase in the tissue column to generate CL signal. The linear range of calibration curve was 1.0 x 10(-6) similar to 1.0 x 10(-3) mol/L and the determination limit was 4 x 10(-7) mol/L. The biosensor was stable for over 400 determinations of hydrogen peroxide during a period of 3 weeks. A complete analysis, including sampling and washing, could be performed in 1.5 min with a relative standard deviation of 2.4% The system has been successfully applied to the determination of hydrogen peroxide in water samples.