Abstract:The chemiluminescence of periodate and hydrogen peroxide in KOH–K2CO3 medium has been developed as a flow-through sensor for the determination of H2O2 in the rainwater. Periodate, hydroxyl and carbonate ions were immobilized on a strongly basic anion-exchange resin. Several different resins, e.g. Amberlite IRA series, Amberlyst A-26 and A-27 and Muromac MSA-1 and SBR were compared. The preparations of the resins in IO4−, OH− and CO32− forms were also discussed. Using the mixture of two different IRA-458 resins separately treated with 0.02 M KIO4 solution and 0.9 mol l−1 KOH–0.4 mol l−1 K2CO3 solution, the sensor can be used to determine H2O2 in the range of 2.0×10−7–1×10−4 mol l−1. The limit of detection is 1.0×10−7 mol l−1 H2O2 and the relative standard derivation is 3.8% for the 1×10−6 mol l−1 H2O2 solution in 15 repeated measurements. This method has been successfully applied to determine H2O2 in rainwater without any sample pre-treatment. The recoveries for all samples were 90–107%.