An Unmediated Hydrogen Peroxide Sensor Based on A Hemoglobin-Sds Film Modified Electrode

CH Fan,I Suzuki,Q Chen,GX Li,J Anzai
DOI: https://doi.org/10.1080/00032710008543212
2000-01-01
Analytical Letters
Abstract:An unmediated hydrogen peroxide sensor is designed in this paper by employing a hemoglobin-SDS film modified electrode. Hemoglobin exhibits direct (unmediated) electrochemistry at the modified electrode. The protein also shows elegant catalytic activity towards the electrochemical reduction of hydrogen peroxide. Consequently, a prototype hydrogen peroxide sensor is prepared. Under optimum conditions, this sensor provides a linear response over the hydrogen peroxide concentrations in the range of 1x10(-5)similar to 1x10(-4) mol/L. The detection limit was 2x10(-6) mol/L The relative standard deviation was 4.2% for 6 successive determinations of the hydrogen peroxide at 1x10(-5) mol/L. This configuration is shown to be sensitive, stable and easily fabricated. It might be useful in the biological and industrial fields.