TiO 2 sol-gel derived amperometric biosensor for H 2 O 2 on the electropolymerized phenazine methosulfate modified electrode
Xin Xu,Jiequan Zhao,Dechen Jiang,Jilie Kong,Baohong Liu,Jiaqi Deng
DOI: https://doi.org/10.1007/s00216-002-1616-0
2002-01-01
Analytical and Bioanalytical Chemistry
Abstract:. A novel hydrogen peroxide biosensor was developed based on the immobilization of horseradish peroxidase (HRP) in a TiO 2 sol-gel matrix on an electropolymerized phenazine methosulfate (PMS) modified electrode surface. Such membranes are of interest due to their high surface area, biological compatibility, and ease of fabrication. HRP entrapped in the TiO 2 matix was stable and retained its activity to a large extent. Cyclic voltammetry and amperometric measurements were employed to demonstrate the feasibility of electron transfer between immobilized HRP and the glassy carbon electrode via electropolymerized PMS. The influence of various experimental parameters such as operating potential, pH, temperature, and stability was investigated for optimum analytical performance. The biosensor provided a wide linear calibration range from 4.0×10 –6 M to 1.0×10 –3 M, with a detection limit of 8.0×10 –7 M at a signal-to-noise ratio of 3. The sensor retained 80% of its original activity after two months of operation.